Free Chlorine Analyser M 1036 C

Products

Free residual chlorine analyser series M 1036 C is designed for measurement and control of drinking water, wastewater treatment and swimming-pool applications.

Two basic measured parameters are free chlorine and temperature.

Add to inquiry list

Measuring range 0 – 5 mg/l

KEY FEATURES

  • 3 measuring electrodes platinum/platinum + reference electrode
  • continuous measuring: free chlorine
  • Stable measurement— open flow measurement cell with a constant flow
  • Continuous cleaning of measuring electrodes by stepper motor
  • Insight in measurement cell parameters for rapid detection of interfering substances
  • Modular electronics for simple service
  • Alarm for measuring cell or regulation errors
  • Canbus communication to easily connect several units to the control unit
  • Clear graphical display for all measured parameters

GENERAL

Free residual chlorine analyser series M 1036 C is designed for measurement and control of drinking water, wastewater treatment and swimming-pool applications.

Two basic measured parameters are free chlorine and temperature.

The unit enables the construction of additional expansion modules, for measuring the pH value and the redox potential (for drinking and wastewater).

Amperometric measurement method using two electrodes and an additional third reference electrode provides constant and continuous measurements.

The unit consists of the following subassemblies:

  • a mechanical filter sample
  • an indicator of the flow of the sample
  • self-cleaning measuring cell
  • measurement – regulation electronics
  • cover

The user interface for parameterization and calibration is performed via graphical LCD display.

Control, regulation and control of disinfection processes are enabled by analog and digital outputs and the CANbus digital communication. The control module supports the connection of actuators – dosing pump with analog and digital inputs as well as solenoid and motorized valves – for two-point and continuous dosing.

Software design of the control module contains a number of protective mechanisms against overdosing.

Ambient tempterature
-10...+50C
Relative humidity
10...95% non-condensating
Control unit protection
IP 65
Dimensions
670 x 322 x 113
Weight
6,5kg
Flow
minimum 0,5l/min

Measuring range (M.R.):
0 - 5 mg/l
Measuring principle: (amperometric):
Platinum-platinum electrodes and third reference electrode
Measured value resolution:
0,01 mg/l
Deviation of indication, measured value:
maximum 2% of MR
pH range:
6.5...8
Water conductivity:
> 50µS
Automatic temperature compensation range:
0...+50ºC

Electrical data

Supply voltage:
24 VAC/VDC ±20%
Power:
5 W Maximum
Isolation voltage:
500V (to analogue part)

Current output connector

Modules
2
Current range
4...20mA
Load
maximum 800 E at 24VDC
Output range:
adjustable to (MR)
Output Mode options:
- free chlorine signal output
- temperature
- regulator output

Regulator connector

Regulator
3 point or PI regulator
Outputs
3 x 24 VAC / 250 mA
Inputs
3 x 24 VAC
Mode options
- Motor Control
- Pump Control

Documents

BUL00048.pdf
Add to inquiry list

Interested in a custom tailored solution?

Contact us